Quality Control
For manufacturing of surgical and dental instruments,
MLINTRA first chooses the correct stainless steel grade for required
instruments. MLINTRA takes the help of ASTM & DIN standards, DIN book,
ISO's standard manual and other suitable literature for the manufacturing of
surgical instruments.
For surgical instruments, mostly two types of stainless
steels are used:-
• AUSTINITIC
STAINLESS STEEL (AISI 300 Series) Non-Magnet.
• MARTENSITIC
STAINLESS STEEL (AISI 400 Series) Magnet.
Mostly Martensitic stainless steel is used with AISI 410
& AISI 420 grades for the manufacturing of General Surgical Instruments.
Austenitic and Martensitic stainless steel are used for the
manufacturing of Dental instruments.
AISI 410 Grade is used for manufacturing of NON-CUTTING Instruments.
AISI 420 Grade is used for the manufacturing of CUTTING
Instruments.
AISI 304 is used for the manufacturing of the Tips of
Scalars and Elevator Handles.
Other Stainless steel Grades are also used according to
requirement of the instruments.
Quality Processes
1. Hardness
• Material
shall conform to the mechanical property requirements as following:
Hardness Guidelines for Selected Class 4 Martensitic
Stainless Steels in the Annealed Condition
Type Brinell Hardness
Max
(HB)
410 210
410X 220
416 262
416 Mod 262
420A 220
420B 235
420 Mod 255
420X 262
420C 262
420F 262
420F Mod 262
431 285
440A 285
440B 285
440C 285
440F 285
2. Corrosion Resistance
There are two test methods are specified for determining
corrosion resistance
Copper Sulfate Resistance Test and Boiled Water Resistance
Test.
A-Copper Sulfate Resistance Test
Check that there is No copper deposit on the instruments.
B-Boiled Water Resistance Test
Check that there is No blemishes on the instrument OR No
visible signs of corrosion.
3. Elasticity
A-Function of needle holders
& Haemostatic forceps
The function of needle holders shall be tested in accordance
with the following:
Place a plastics fiber (e.g. a suture filament) of a maximum
diameter 0.2mm between the jaws of the instrument at a point within the third
of the length nearest the tip. Fully close the instrument and apply a tensile
force of 20N to the fiber. Record whether the fiber is pulled out from the
jaws.
After the test, no distortion, cracks or any other permanent
modifications shall be visible.
B-Function of Haemostatic
forceps
Place a test wire in accordance with table 1 or 2 as
appropriate between the tips of the instrument jaws. Fully close the instrument
to the last ratchets position. Leave the instrument in this position for 3 h at
room temperature. Examine the instrument for the presence of cracks and
permanent deformation.
4. Cutting (for cutting
instruments)
The cutting ability of the instruments shall be tested in
accordance with the following:
The testing material in accordance with the table shall be
cut non-stop three times along two-third of the blade without lateral pressure.
• The
testing material shall be cut cleanly without tearing.
• After the
test, no distortion, cracks or any other permanent modifications of the
instrument shall be visible.
5. Visual Inspection
All Instruments should pass the following inspections:
• The
instruments shall have joints which move smoothly and shall be neither too
loose nor too tight; it shall be possible to close and reopen the instruments
easily with two fingers.
• All
surfaces shall be free from pores, crevices and grinding marks. The instrument
shall be supplied free from residual scale, acid, grease and grinding and
polishing materials. Compliance with these requirements shall be checked by
inspection under normal vision, corrected, if necessary.
• Check the
instruments marking with Life Care trade mark & Code of the instruments.
• Instruments
having tungsten carbide inserts shall have gold-colored handles.
• Check the
instruments dimensions according to Catalogue.
6. Storage and sterilization
Be sure that warranty & instructions are properly
sterilized.
7. Packaging
• The
packing design should be fulfill protected of each instrument against damage
and deterioration.